Revision as of 18:03, 12 June 2024 by Jblocher (talk | contribs) (Created page with "== Acquisition topics == {| | # Comparison 4 parallel coil and standard RF head coil: #* FA and MD comparable within less than 5-10% #* RF shading different in anterior/frontal area #* Geometry distortions/differences in both anterior and posterior area, most in posterior superior area #* Tissue intensities: about 5%-6% gain in four channel coil #** 4C: WM/GM temporal 441/291 = 1.51; central (caudateGM) 515/435 = 1.18 #** Standard: WM/GM temporal 399/278 = 1.43; centr...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search

Acquisition topics

  1. Comparison 4 parallel coil and standard RF head coil:
    • FA and MD comparable within less than 5-10%
    • RF shading different in anterior/frontal area
    • Geometry distortions/differences in both anterior and posterior area, most in posterior superior area
    • Tissue intensities: about 5%-6% gain in four channel coil
      • 4C: WM/GM temporal 441/291 = 1.51; central (caudateGM) 515/435 = 1.18
      • Standard: WM/GM temporal 399/278 = 1.43; central (caudateGM) 371/415 = 1.11
File:DifferenceImageCoilComparison.jpg
Difference image shows both the shading differences and the geometry distortion
File:OverlayCoilComparison.jpg
Red: 4 grad coil, Green: standard Head coil. Geometric differences at posterior superior are quite large